Noah Menard, Laboratory Manager forVeritasTesting and Consulting, talks to AZoM about the new TGA-IST16-GC/MS system from Mettler Toledo and how it compares to other options for evolved gas analysis.
What are some of the key features you look for in a TGA?
The most important part of a TGA is the balance. This needs to be highly accurate and sensitive to identify alterations in mass as the entire point of a TGA is to identify weight changes. By accurately determining the change in weight, a user can then have a good idea of the composition of the sample and how each part is affected by the change in temperature or time. To be sure of what those components are, we need to use an EGA technique and work backwards from the offgas. Many evolved gas applications look at components that are coming off at a certain temperature.
In addition to an accurate and high level of sensitivity, the ability to hold a large sample amount is also a significant factor. As with all of the evolved gas analysis techniques, the more sample you are able to load into the TGA results in stronger signals for the corresponding hyphenated technique, especially when looking at events that are a fraction of a percent. Gas flow through the furnace and temperature control are also important.
What is a Hyphenated Technique
A hyphenated technique involves the coupling of one or more instruments together in order to obtain a better picture of the material properties. It gets its name from the hyphen used in designating it – for example TG-IR for TGA coupled to FTIR. In some pharmaceutical areas, this can also be called an orthogonal technique. The majority of these techniques generally involve some form of intermediate interface to regulate material transfer; these can be simple heated lines up to highly precise systems with regulated gas flow, temperature and sampling devices. Evolved gas analysis, which looks at the outgassing components of material, is a prime example and involves methods such as TGA-GC/MS, TGA-MS, and TGA-IR. Other techniques such as Photo-DSC, RAMAN-DSC, UV-DMA, or even GC/MS are also considered hyphenated techniques.
How does TGA-GCMS compare with TGA-IR and TGA-MS? Why would you choose one technique over the other?
There are different advantages to each of the evolved gas techniques. The three most common involve connecting to a FTIR, a MS, and/or a GCMS. TGA-FTIR allows for the analysis of functional groups and isomers of molecules, but is somewhat limited for a few reasons. The wavelengths measured by a standard IR may not be enough for an adequate identification. Another issue is the possible masking of components when CO2and H2O comes off of the sample within the TGA. Out of the EGA techniques, IR is the least sensitive and requires more sample due to the lower detection limits of IR. While IR vapor libraries exist, they are not as extensive as those of a GC/MS. This is probably the most common technique due to simplicity and cost.
TGA-MSallows you to look at mass ions and fragments. This can be very useful when you have an idea of what you are looking for or for simple outgassing issues, but can become difficult when there are multiple overlapping events or a large amount of sample released at one time. Libraries exist but are somewhat limited. It is certainly more sensitive than the FTIR method, but you have to be concerned about the amu limit of the instrument and flooding the sensor. This is a great technique for gases like CO2, water, and very light materials.
分析商用杏仁味电子烟溶液,作为TGA-GC/MS鉴定复杂矩阵中组件的能力的一个例子。将调味剂和尼古丁的丙烯甘油/甘油溶液分开,是一个良好的模型系统,适合困难样品。
GC/MS alone is one of the most sensitive and commonly used techniques due to its ability to separate out components from complex materials. This is what makes it a workhorse in analytical laboratories. When paired with a TGA, you are able to then analyze each weight loss event while being able to prevent sample overload by different injection methods, as well as taking advantage of the high instrument sensitivity. Each segment can be analyzed individually for a further understanding of the material composition. Since GC/MS is used so universally, there are extensive libraries that can identify each component with little user interaction. When you are looking at a fraction of a percent weight loss in a TGA, a GC/MS coupling excels as it can easily detect and separate the released gas.
How does the new TGA/DSC3+ - IST16 - GC/MS compare to other options for evolved gas analysis (such as TGA-FTIR and TGA-MS)?
TGA-IR和TGA-MS都不采样,而是直接从TGA运行进化的气体混合物。只有当您开始使用色谱技术时,采样才成为一种选择,但是较旧的方法每次运行仅使用1或2个样本。由于TGA-GC/MS的强度是分离组件的能力,因此更多样品有助于解决混合物。Mettler-Toledo系统,该系统伴侣TGA/DSC3 to an IST 16 to a GCMSallows collecting of 16 independent samples from the TGA run. On a normal run where we are looking for components under 600 C, we can sample every 40 degrees. Of course, we can also concentrate our sampling in a region of interest to get even more information.
While TGA-IR and TGA-MS do allow for real time scanning, it is a continual stream of gas flow which does not allow targeted sampling. As there may be large amounts of molecular fragments being transferred, masking of the signal due to the earlier evolved gas flooding the system may occur. With the IST system taking designated sampling locations, this problem can be averted. This also helps with liquids and other similar materials that greatly expand when evaporation, as the sample loop holds a small amount; when run through a GC-MS split/splitless mode the risk of flooding of the detectors becomes even easier to avoid.
位于Mettler Toledo TGA/DSC3+和Agilent GC/MS之间的SRA存储界面的独特之处是什么?
The first unique capability is the ability to store up to 16 gas samples in the internal loops. Many of these tests have more than one location of interest, so it fits well with most laboratory requirements. Some models that facilitate evolved gas analysis testing run a continual stream of gas from the TGA, which can cause the user to need to repeat the run more than once to obtain all the data required for a full analysis. At this point, it may become harder to identify which component is coming off at a location when there are multiple weight loss events occurring. As smaller molecules will move through the GC/MS faster, a cluster of higher mass molecules may form giving an inexact idea of what came off at each location.
此外,这些可以将TGA绑起来,直到整个过程完成并需要连续的用户互动,从而将人员联系在一起并减少实验室吞吐量。另一个优点是在TGA运行后,IST系统可以与TGA断开连接。IST系统可以协调样品注射,并让气体分析自动继续,而无需任何进一步的交互作用,直到分析所有样本循环,而TGA则可以用于其他项目。根据样品要求,可以通过软件控制传输线温度和储物室温度。此外,用户有能力通过凝视软件制作集合循环触发,或者通过输入何时进行样品来通过界面进行样品。
在最多16个循环中存储进化的气体的优势是什么?
TheIST-16 systemallows us to take several measurements during the course of a TGA run. For some materials there may be several weight loss events during the course of heating. With sixteen loops we can take samples from all the different locations of interest in a single run, allowing greater confidence in the results as well as determine how the molecule fragments during decomposition. For an example, taking several samples during a single weight loss event can show what fragments come off first and the decomposition process based on the progression through the weight loss.
存储步骤期间有样本转换的风险吗?
大多数变化发生在材料的分解或超出材料时的TGA中。如果是溶剂,则系统将很容易地存储样品而无需更改。当您开始分解样本时,片段通常会在已经稳定的离散段中脱落。即使这些碎片会随着时间的流逝而衰减,循环中的时间也很可能足以引起重大变化。
是否可以用IST16(BP> 200°C)分析高沸点化合物?
This is not a problem in the system, as the heated chamber allows the sample to remain in gas form while awaiting injection into the GC. The GC parameters can be optimized for that analysis while the IST loops hold the samples.
Is it possible to perform more than 16 injections in the GC column from one TGA experiment?
Well, it depends. One approach is to use the multi-injection mode, which can be used to collect and inject samples in a short amount of time during the TGA run. This sends pulses of gas into the GC at a user set rate and period of time. There is a slight loss in precision of result, but it is offset by the ability to take several more samples than normal.
Can you tell us about a success story where the instrument has been used to solve a particularly difficult problem?
While there are several, two stand out. One application was for an issue with piping in a chemical plant. Certain pipes were turning brittle and cracking after a certain time within the plant. This was localized to only one plant, and the problem only appeared intermittently. While DSC testing indicated some difference between fresh material, known good material, and corrupted material replication of the problem proved elusive. We were able to rule out thermally induced degradation, and evidence pointed towards chemical degradation. As it was a chemical plant, we began to wonder if there were leaks or release of solvent vapours as direct sample soaking in solvents appeared to degrade the material well past the actual degraded sample. Since these problems took several months to years to appear, there was no real way to directly measure what occurred. In this case, we were able to heat the tubing to degradation and analyze the gas composition. From this we were able to determine that a solvent by-product produced in the plant was being released into the atmosphere in small amounts intermittently, and collecting in the tubing which in turn caused the cracking and degradation observed.
IST16原理
Another example was in pharmaceutical manufacturing. In many cases, there are several solvents used in the creation of medicines, and improper removal can cause residual solvents to remain in the final product. This is obviously a sub-optimal situation for the manufacturer as the FDA takes a dim view of this. We found that when heated in the TGA, we could identify not one, but four solvents associated with the weight loss. From that, we could tell the client where in the manufacturing process the issue was occurring. Other techniques may miss these because of the low levels of each solvent or because of overlapping fragmentation patterns in TG-MS.
了解请求数据的客户的解决方案的价值是什么?
最重要的是,它表明工厂内的工人可能存在健康和安全问题。此外,这表明过程必须重新评估,因为它正在引起植物基础设施问题。尽管对问题的识别具有一些短期收益,但客户能够实施可以防止未来问题的变更,随着时间的流逝,这可能会逐渐变得更糟。由于这些是高通量植物,因此即使是部分或临时关闭也可能引起问题。
How easy is the system to operate for someone who is not an expert in chromatograph and thermal analysis?
All of the systems are user friendly, and tie together very well due to the level of automation. From the TGA software we can designate time based locations for sampling, which then sends a signal to the IST system to take a sample at the chosen time. From that, the IST system ties automatically into a user defined GC sequence based on the number of active sample loops. The IST system sends a signal to the GC to initiate the method; when the method is done and reset the IST-16 will then trigger the next sequence until all of the loops have been discharged and analysed. Out of the entire process the hardest step may be determining your GC method for proper compound separation. However, in many cases, a general temperature ramp in the GC is sufficient. With the continued development in modern GC-MS software, this is also easy to apply to the desired amount of loops.
About Noah Menard
Noah has over a decade of experience in hyphenated techniques and thermal analysis.
He is the laboratory manager forVeritas Testing and Consulting, where he oversees lab operations.
In addition to TGA-GC/MS, Veritas Testing and Consulting has a range of thermal and rheological equipment (DSC, TGA, TMA, and DMA) along with other hyphenated techniques such as TGA-IR, Photo-DSC, and Photo-DMA.
免责声明:此处表达的观点是受访者的观点,不一定代表本网站的所有者和运营商Azom.com Limited(T/A)Azonetwork的观点。此免责声明构成了Terms and Conditionsof use of this website.