How to Conduct a Tensile Test of CNF-Reinforced Plastic at High-Speed

Cellulose is well known as the principal component of plants. It is also the most plentiful carbohydrate in the world, used as a raw material in textile and paper manufacturing since antiquity. In more recent years, cellulose nanofiber (CNF), with higher functionality achieved by defibrating cellulose to the nano level, has attracted significant attention.

CNF has minimal environmental impacts because it is a plant-derived material. It also possesses a wide range of desirable functions such as a gas barrier property, low linear expansion, and transparency. CNF is lightweight - weighing only a fifth as much as steel - while exhibiting high specific strength between 5 and 8 times stronger than ferrous materials.

The research aimed at the development of composite materials with a combination of high strength and lightweight is currently in progress. Researchers hope that by adding CNF to thermoplastic resins, these new materials will be beneficial in a range of sectors.

One such sector is the automotive industry, where a key goal remains the improvement of fuel efficiency via a reduction in auto body weight. Considering the use conditions of automobiles, it is important that clarity is achieved not only in relation to the static mechanical properties of materials, but also of their fatigue characteristics, impact properties, and thermal characteristics.

This article sees the tensile strength of CNF-reinforced plastic being evaluated at numerous test speeds (strain rates) using a high-speed impact testing machine and a Shimadzu testing machine. Following these tests, the fracture surfaces of the specimens were also observed using an electron probe microanalyzer (EPMA).1

Measurement System

Figure 1 displays the condition of the high-speed tensile test and the static tensile test. AnAGS-X table-top type universal testing machinewas employed in low-speed tests, while a HITS™-TX high-speed impact testing machine was employed in high-speed tests.

Condition of Tests. (a) Static Tensile Test, (b) High-Speed Tensile Test.

Figure 1.Condition of Tests. (a) Static Tensile Test, (b) High-Speed Tensile Test.

An optical microscope and an EPMA™- 8050G electron probe microanalyzer were utilized in observations of the specimens’ fracture surfaces after the tests. Table 1 illustrates the range of test equipment utilized in this study.

Table 1.Test Equipment.

. . .
Testing machines : AGS-X (static tensile tests)
HITS-TX (high-speed tensile tests)
Load cell : 1 kN (AGS-X)
2 kN (HITS-TX)
Grip : Pneumatic flat grip (AGS-X)
Grip for plate specimens (HITS-TX)
Fracture surface observation : EPMA-8050G

Table 2.Test Conditions and Specimen Information.

. . .
Test speed : 0.0001, 0.001, 0.01, 0.1 /s (AGS-X)
0.1, 1, 10, 100 /s (HITS-TX)
(0.000004 - 4 m/s)
Test temperature : Room temperature
Number of tests : n = 3
Specimens : HDPE, CNF10%/HDPE
Specimen dimensions : T1 mm × W5 mm, parallel part 40 mm

Measurement Results

Two types of specimens were prepared. One type used high-density polyethylene (HDPE) while the other used HDPE reinforced with 10% CNF (CNF10%/HDPE). Table 2 displays the full test conditions and related specimen information.

Stress-Displacement Curves. (a) HDPE, (b) CNF10%/HDPE.

Figure 2.Stress-Displacement Curves. (a) HDPE, (b) CNF10%/HDPE.

Figure 2 gives examples of the stress-displacement curves of both the HDPE and CNF10%/HDPE at different strain rates. As can be seen in Figure 2, both materials demonstrated lower tensile strength as the test speed decreased while demonstrating higher tensile strength at higher test speeds. Additionally, the slopes of the stress-displacement curves were also seen to become larger as test speed increased.

Relationship Between Strain Rate and Tensile Strength.

Figure 3.Relationship Between Strain Rate and Tensile Strength.

Figure 3 displays the relationship between the tensile strength and strain rate. The results in Figure 3 show that CNF10%/HDPE exhibited larger tensile strength under every strain rate condition. The tensile strength was also found to increase at higher test speeds.

Images of Lateral Surfaces of Specimens after Test. (a) CNF10%/HDPE, 100 /s, (b) CNF10%/HDPE, 0.0001 /s, (c) HDPE, 100 /s, (d) HDPE, 0.0001 /s.

Figure 4.Images of Lateral Surfaces of Specimens after Test. (a) CNF10%/HDPE, 100 /s, (b) CNF10%/HDPE, 0.0001 /s, (c) HDPE, 100 /s, (d) HDPE, 0.0001 /s.

Figure 4 displays images of the specimens’ lateral surfaces after the test was photographed with the optical microscope. The left side of the images shows the fracture surface. Figures 4(b), (c), and (d) indicate that whitening has taken place around the fracture surfaces.

These images indicate that the fracture mode is a ductile fracture. In Figure 4(b), a pale diagonal stripe has appeared, but, because Figure 4(a) reveals practically no traces of whitening, it is believed that brittle fracture took place in this specimen.

Images of Fracture Surfaces of Specimens after Test. (a) CNF10%/HDPE, 100 /s, (b) CNF10%/HDPE, 0.0001 /s, (c) HDPE, 100 /s, (d) HDPE, 0.0001 /s.

Figure 5.Images of Fracture Surfaces of Specimens after Test. (a) CNF10%/HDPE, 100 /s, (b) CNF10%/HDPE, 0.0001 /s, (c) HDPE, 100 /s, (d) HDPE, 0.0001 /s.

Figure 5 displays images of the fracture surfaces after the test was photographed using the optical microscope. Figure 6 reveals the EPMA images of the parts defined by the white frames in Figure 5. Meanwhile, Figure 6(a) shows a condition whereby fractures had occurred in fibers in which defibration had not progressed.

EPMA Observation Images of Fracture Surfaces. (a) CNF10%/HDPE, 100 /s, (b) CNF10%/HDPE, 0.0001 /s, (c) HDPE, 100 /s, (d) HDPE, 0.0001 /s.

Figure 6.EPMA Observation Images of Fracture Surfaces. (a) CNF10%/HDPE, 100 /s, (b) CNF10%/HDPE, 0.0001 /s, (c) HDPE, 100 /s, (d) HDPE, 0.0001 /s.

Fracture of the fibers cannot be seen in Figure 6(b), but a condition of ductile elongation was observed in the resin which contained CNF. The image in Figure 6(c) is the result of an observation of the most elongated part seen in Figure 5. Ductile elongation of the resin was exhibited in this figure, while the elongation of the resin could also be seen in Figure 6(d).

Conclusion

In this study, a high-speed tensile test of CNF10%/HDPE and HDPE was performed. The results confirmed that tensile strength altered depending on test speed. Additionally, fracture surfaces after the tests were observed, explaining the fact that differing fracture surfaces had been found.

TheHITS-TXand AGS-X allow a wide range of strain rates to be defined as test conditions, meaning that these testing machines are highly suited for evaluation of the strain rate dependency of resins.

References and Further Reading

  1. Fumiaki Yano, Yuki Kamei, Takeshi Senba, and Kazuo Kitagawa, Abstracts of the Symposium on Polymer Processing ’19 (2019)

This information has been sourced, reviewed and adapted from materials provided by Shimadzu Scientific Instruments.

For more information on this source, please visit日本岛津公司的科学struments.

Citations

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    日本岛津公司的科学struments. (2020, October 26). How to Conduct a Tensile Test of CNF-Reinforced Plastic at High-Speed. AZoM. Retrieved on July 06, 2021 from //www.wireless-io.com/article.aspx?ArticleID=19522.

  • MLA

    日本岛津公司的科学struments. "How to Conduct a Tensile Test of CNF-Reinforced Plastic at High-Speed".AZoM. 06 July 2021. .

  • Chicago

    日本岛津公司的科学struments. "How to Conduct a Tensile Test of CNF-Reinforced Plastic at High-Speed". AZoM. //www.wireless-io.com/article.aspx?ArticleID=19522. (accessed July 06, 2021).

  • Harvard

    日本岛津公司的科学struments. 2020.How to Conduct a Tensile Test of CNF-Reinforced Plastic at High-Speed. AZoM, viewed 06 July 2021, //www.wireless-io.com/article.aspx?ArticleID=19522.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Submit